Handpiece Bearing Care

 1         Clean external surfaces of handpiece.

2         Insert bur into chuck.

3         Clean handpiece internally as follows.

Spray handpiece cleaner or cleaner/lubricant into drive air turbine for 1 to 2 seconds.

Operate handpiece within a paper towel or other cover to purge used lubricant and debris.

Repeat the spray/operation procedure until expelled fluid is absolutely clear.

Wipe external surfaces to remove expelled fluid.

            Operate handpiece for 3-5 seconds, repeating until all fluid has been expelled.

 

4         Sterilize handpiece according to manufacturer's instruction.

5         Cool handpiece to room temperature.

6         Apply two drops of lubricant into drive air tube.

7         Operate handpiece for 5 to 10 seconds within paper towel or other cover to purge excess lubricant.

8         Wipe external surfaces.

9         Handpiece is ready for use.


  • Keine Produktkommentare vorhanden
E-Mail-Adresse
Wertungen
Produktbewertung
Customer Service
Bezahlung
Lieferung & Rücksendung
Verbinden uns
Company Info
Geheimnis & Sicherheitspolitik
Über Athenadental.de
Wer sind Wir?
NUTZUNGSBEDINGUNGEN
RECHTE AN GEISTIGEM EIGENTUM
Brauchen Sie Hilfe?
Über unsere Website
Über Ihr Paket
Newsletter anmelden & Vorteile sichern

E-Mail-Adresse: service@athenadental.de

Bitte geben Sie eine gültige E-Mail-Adresse an
close

HERZLICHEN DANK für Ihr Interesse an unserem AthenaDental-Newsletter.

Sie erhalten von uns in Kürze eine E-Mail mit einem Link, den Sie bitte als Bestätigung Ihrer Anmeldung anklicken.

Erst nach dem Klick ist Ihre Anmeldung mit der E-Mail-Adresse:  aktiv.

Wir freuen uns auf Sie!

athenadental.de

Copyright service@athenadental.de ©2020 Alle Rechte vorbehalten.